Common questions

What should a dental sterilisation log record?

The date, the cycle number, the steriliser, what was in the load, whether the cycle reached its parameters, the chemical indicator result, any test result, and who released the load. That lets you trace an instrument back to its cycle.

How often should a dental steriliser be spore tested?

Many guidelines call for a biological (spore) test at least weekly, and with every load that contains an implant. For example, the US CDC recommends at least weekly. Check your own guidelines and the maker’s instructions.

What’s the difference between chemical and biological indicators?

A chemical indicator changes colour when the pack has been exposed to the right conditions, and is checked on every pack. A biological indicator contains bacterial spores, and shows whether the cycle actually killed them. It’s used for periodic tests.

What should happen when a steriliser cycle fails?

Don’t use the load. Take the steriliser out of use, reprocess the instruments in a working steriliser, record the failure, and find the cause before using it again. A failed spore test may mean recalling instruments from earlier loads, so follow your failure procedure.

Can Wikidoc record the steriliser checks?

Yes. Open the log in Wikidoc as a daily checklist. Whoever runs the steriliser marks the daily test and the day’s loads pass or fail, and records the test cycle’s temperature against the range your steriliser’s maker gives. A failed check needs a note on what was done before the day can be signed off. Keep the printouts or the paper log as well if your rules ask for them.

Run your checklists and SOPs in one place